| Model Number |
SRLP-235 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
03/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture, baker grade iii" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare provider reported left side capsular contracture, baker grade iii.The device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare provider reported left side capsular contracture, baker grade iii.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified: opening curved on anterior and weight to the spec.A visual and microscopic analysis was performed which identified: opening punctured on anterior, non-penetrating nicks, wear abrasion and creases fold.Based on the device analysis the final assessment is: a punctured on anterior assessed as surgical damage like.
|
| |
|