| Model Number |
110-360 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Failure of Implant (1924)
|
| Date of Event |
03/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Patient reported left side rupture.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The event of capsular contracture, baker grade iii/iv is a physiological complication.And analysis of the device generally does not assist allergan in determining a probable cause for this event.
|
| |
|
Event or Problem Description
|
|
Physician reported, capsular contracture, baker grade iii/iv.And concerns with the product.Device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Patient reported left side rupture.Physician reported capsular contracture, baker grade iii/iv and concerns with the product.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: "based on the device analysis grid, the assessments of the complaint are: capsular contracture: unable to observe as it is a medical event not related to the device.Rupture: no observed rupture on the device.Anxiety-product/procedure: unable to observe as it is a medical event not related to the device.Additional observations: crease fold observed on the device.Deformation device observed on the device.Yellow particles observed on the device.No further actions are required as the device was implanted." additional, changed, and/or corrected data.
|
| |
|