PENUMBRA, INC. PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER; NRY Back to Search Results
Catalog Number 5MAXACE068
Medical Device Problem Codes Break (1069); Material Deformation (2976); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the left middle cerebral artery (mca) using a penumbra system ace 68 reperfusion catheter (ace68) and a non-penumbra sheath.During the procedure, the physician experienced resistance while advancing the ace68 through the sheath.Subsequently, the physician kinked and broke the ace68 on the proximal end, five centimeters from the hub.Therefore, the ace68 was removed.The procedure was completed using a new ace68 and the same non-penumbra sheath.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Additional Manufacturer Narrative
Evaluation of the returned ace68 confirmed a fracture on the proximal shaft.If the device is manipulated at extreme angles during advancement against resistance, damage such as a kink and subsequent fracture may occur.The unraveled coil winds likely occurred during removal of the device from the procedure.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key14881109
Report Number3005168196-2022-00316
Device Sequence Number6245159
Product Code NRY
UDI-Device Identifier00814548016535
UDI-Public814548016535
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K161640
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/20/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/24/2024
Device Catalogue Number5MAXACE068
Device Lot NumberH11509
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2022
Supplement Date Received by Manufacturer07/11/2022
Initial Report FDA Received Date06/30/2022
Supplement Report FDA Received Date07/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexFemale
Date Report Sent to FDA01/01/2005
-
-