MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® DUALMESH® PLUS BIOMATERIAL; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® DUALMESH® PLUS BIOMATERIAL; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number 1DLMCP04
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Unspecified Infection (1930); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The plus antimicrobial product coating contains silver carbonate, approximately 800 micrograms per cubic centimeter of product (g/cm3), and chlorhexidine diacetate approximately 1600 micrograms per cubic centimeter of product (g/cm3).It should be noted that the gore® dualmesh® plus biomaterial instructions for use addresses the following adverse reactions among others: ¿possible adverse reactions with the use of any tissue deficiency prosthesis may include, but are not limited to, contamination, infection, inflammation, adhesion, fistula formation, seroma formation, hematoma, and recurrence.The gore® dualmesh® plus biomaterial instructions for use also states: ¿strict aseptic techniques should be followed.If an infection develops, it should be treated aggressively.An unresolved infection may require removal of the material.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
It was reported to gore that the patient underwent laparoscopic ventral hernia repair on (b)(6) 2002 whereby a gore® dualmesh® plus biomaterial was implanted.The complaint alleges that on (b)(6) 2016, an additional procedure occurred whereby the gore device was explanted.It was reported the patient alleges the following injuries: additional surgery, recurrence, diarrhea, infection, leukocytosis, nausea, pain & suffering.Additional event specific information was not provided.
 
Additional Manufacturer Narrative
H6: conclusion code remains unchanged.It should be noted that the gore® dualmesh® plus biomaterial instructions for use addresses the following adverse reactions among others: ¿possible adverse reactions with the use of any tissue deficiency prosthesis may include, but are not limited to, contamination, infection, inflammation, adhesion, fistula formation, seroma formation, hematoma, and recurrence.The gore® dualmesh® plus biomaterial instructions for use also states: ¿strict aseptic techniques should be followed.If an infection develops, it should be treated aggressively.An unresolved infection may require removal of the material.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
B7: added medical history.H6: conclusion code remains unchanged.H10/11: added medical record information.Additional details regarding the patient's clinical course were ascertained from a review of medical records and are as follows: implant procedure: laparoscopic ventral hernia repair.Implant: gore® dualmesh® plus biomaterial w/ corduroy surface ref: 1dlmcp04 ; lot: 01269508 implant date: (b)(6), 2002 [hospitalization (b)(6), 2002] ¿ (b)(6) 2002: (b)(6) hospital (b)(6) (b)(6), va).(b)(6), m.D.Operative report.Pre- and postoperative diagnosis: ventral hernia.Anesthesia: general.Estimated blood loss: minimal.Specimens: n/a.¿ indications: ¿this patient is a 27-year-old female who is status post procedure with an incision hernia at the periumbilical position.She was brought to the operating room for laparoscopic ventral hernia repair.¿ ¿ procedure: "after obtaining an informed consent, the patient was brought to the operating room and placed supine on the operating room table.After induction of general endotracheal anesthesia, the patient was prepped and draped in the usual sterile fashion.Access was gained to the peritoneal cavity through a right upper quadrant incision using the hasson technique.Once the hasson cannula was secured in place, the abdomen was insufflated with 15 mmhg.A 360 degree look around revealed there to be adhesions in the superior portion of the incision.We took these adhesions gently.Only a couple of adhesions needed to be lysed in order to have circumferential access around the hernia.There were no incarcerated hernia contents.There were two hernias side by side, and the defects collectively measured a 10 cm circle.A 14 cm piece of dual mesh was placed after #0 gore suture was secured in four locations evenly spaced around the circumference of the mesh.The mesh was secured using the suture passer device, circumferentially and with transfascial sutures.Between the sutures, the endotacking device was used to secure the mesh edges along the anterior abdominal wall.The mesh easily fit around the perimeter of the hernia defect.The hernia defect was completely covered.At the end of the procedure a 360 degree look around revealed there to be no abnormalities and good hemostasis.Each of the trocars was removed under direct vision.The right upper quadrant hasson cannula was closed with 1-0 vicryl figure-of-eight suture.The wound was irrigated copiously with antibiotic solution.The skin was closed with 4-0 vicryl suture.Each of the four suture sites was closed and approximated with steri-strips and benzoin.A sterile dressing was placed on each of the incisions.A total of three trocars were used; a 10 mm and two 5 mm trocars on the right side of the abdomen.The patient tolerated the procedure well and was transferred to the recovery room with stable vital signs." ¿ (b)(6) 2002: (b)(6) hospital (b)(6) ((b)(6), va).Implant information: gore® dualmesh corduroy antimicrobial.Item: 1dlmcp04; lot: 01269508, size: 15 cm x 19 cm, expiration date: n/a, quantity: (b)(4).¿ the records confirm gore® dualmesh® plus biomaterial w/ corduroy surface (ref: 1dlmcp04; lot: 01269508) was implanted during the procedure.Relevant medical information: no interim medical records.Revision procedure: repair of lower two ventral hernias with 6.4 cm ventral patch.Repair of the upper ventral hernia with 4.3 cm ventral patch.Revision date: (b)(6), 2016 [hospitalization: (b)(6), 2016] ¿ (b)(6) 2016: (b)(6) hospital (b)(6) ((b)(6), va).(b)(6), m.D.Operative report.Preoperative diagnosis: epigastric ventral hernias x 3.Postoperative diagnosis: epigastric ventral hernias x 3 with incarcerated fat.Anesthesia: general.Estimated blood loss: 10 ml.Specimens: n/a.Drains: none.¿ indications: "¿ 41-year-old female with a pmh [past medical history] of chronic pancreatitis s/p distal pancreatectomy and splenectomy 17 years ago.She has chronic abdominal pain and diarrhea which were worsening for the past month.She had a ct scan a week ago and it demonstrated 3 epigastric ventral hernias.I counseled her at length that her worsening abdominal pain and diarrhea might not be secondary to her hernias.She understood but would like to have them repaired.I recommended open repair due to her multiple abdominal surgery.I would combined the inferior 2 hernias into one and repair it with mesh.The superior one will be repaired with mesh.Benefit and risks were discussed with her, consent was obtained." ¿ findings: ¿three epigastric ventral hernias with likely incarcerated preperitoneal fat.The location and size of the defect listed as below.The lowest hernia: 1 cm x 2 cm (transverse) fascial defect.14 cm superior to the umbilicus and 4 cm superior to the upper edge of the old mesh.The middle hernia: 2 cm x 3.5 cm (transverse) fascial defect with a 1.5 cm width fascial bridge from the middle hernia, 15 cm superior to the umbilicus.The highest hernia: 2 cm x 3 cm (transverse) fascial defect with a 4 cm fascial bridge from the middle one, 20 cm superior to the umbilicus.¿ ¿ procedure: "¿a vertical midline incision was made along the scar over the middle ventral hernia using scalpel.The incision was brought down to the fascia using bovie.Of note, her abdominal wall fat was 6 cm deep.The middle ventral hernia was exposed and found contained fat.After the fascial edge was cleared, the preperitoneal fat was completely reduced.The hernia defect was 2.0 x 3.5 cm.The undersurface of the fascia around the fascial edge was dissected with blunt dissection and electrocautery between the parietal peritoneum and the fascia to a depth of 2 cm in order to make a landing zone for the mesh.The lowest hernia defect was then palpated through the middle fascial defect.The lowest hernia contained fat as well and was similarly isolated and reduced.The undersurface of the fascial edge was created to a depth of 2-3 cm.Then the incision was extended cephalad, the upper hernia was found, isolated and reduced.Similarly a 3-cm depth undersurface along the fascial edge was created with blunt and bovie dissection.At this point, a 6.4 cm ventral patch was brought the field.Cardial sutures with #0 ethibond were placed through the fascia and mesh bilaterally and superior to the middle fascia edge and inferior to the lowest fascia edge.The mesh was then inserted through the middle fascial defect and placed against undersurface of the fascia.The sutures were then tied down over the fascia.The mesh was brought up to the fascial edge by pulling the tabs and it was found to lay flat.The lowest fascial defect was closed transversely over and through the mesh using two #0 ethibond sutures in figure of eight.The middle fascial defect was similarly closed transversely over and through the mesh tabs using two #0 ethibond sutures in figure of eight.At this point, a 4.3 cm ventral patch was brought to the field.Cardial sutures with #0 ethibond were placed through the fascia and mesh bilaterally, superior and inferior around the upper hernia fascia edge.The mesh was then inserted through the upper hernia fascial defect and placed against undersurface of the fascia.The sutures were then tied down over the fascia.The mesh was brought up to the fascial edge by pulling the tabs and it was found to lay flat.The upper mesh was overlapped with lower mesh by 1 cm.Then, the upper hernia fascial defect was closed transversely over and through the mesh tabs using two #0 ethibond sutures in figure of eight.The wound was irrigated.Hemostasis was achieved using bovie.The deep subcutaneous layer was closed using interrupted 3-0 vicryl.A #4-0 vicryl was then used in a running subcuticular fashion to close the skin.The skin was dressed with dermabond, and the patient was awoken without complication." ¿ (b)(6) 2016: (b)(6) hospital (b)(6) ((b)(6), va).Implant information.Ventral patch 6.4 cm.Ventral patch 4.3 cm.Quantity: (b)(4).¿ (b)(6) 2016: (b)(6) hospital (b)(6) ((b)(6), va).Pathology report.Report not provided.A potential relationship, if any, between the alleged injuries or complications and the gore device has not been established at this time based on available information.It should be noted that the gore® dualmesh® plus biomaterial instructions for use addresses the following adverse reactions among others: ¿possible adverse reactions with the use of any tissue deficiency prosthesis may include, but are not limited to, contamination, infection, inflammation, adhesion, fistula formation, seroma formation, hematoma, and recurrence.The gore® dualmesh® plus biomaterial instructions for use also states: ¿strict aseptic techniques should be followed.If an infection develops, it should be treated aggressively.An unresolved infection may require removal of the material.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
The investigation has been completed.Based upon the totality of the information received over the course of the investigation and reported by gore in the previously submitted medical record narratives the following conclusions have been reached.As previously reported, all pertinent medical records requested may not have been received.Through gore's investigation and based on the available information there is no available information that reasonably suggests that a gore device may have caused or contributed to death, serious injury or reportable malfunction, and is no longer considered reportable.Therefore, this event is being coded as no clinical signs, symptoms or conditions, no health consequences or impact and will be closed as no problem detected.Previous patient codes were reported based on the original complaint and are no longer applicable and/or not reportable per gore¿s investigation.Following gore¿s investigation, the previously submitted annex f code has been updated to reflect gore¿s final conclusion.It should be noted that the gore® dualmesh® plus biomaterial instructions for use include warnings and addresses the following adverse reactions among others: ¿possible adverse reactions with the use of any tissue deficiency prosthesis may include, but are not limited to, contamination, infection, inflammation, adhesion, fistula formation, seroma formation, hematoma and recurrence.¿ the instructions for use further warn: ¿as with any implantable device, strict aseptic techniques should be followed.If an infection develops, it should be treated aggressively.An unresolved infection may require removal of the device.¿ procedure and specific patient factors may contribute to or cause infection, leading to contamination, exposure, lack of incorporation and/or seeding of device.Procedure related factors may include adherence to clinical guidelines on infection risk management, contamination of device prior to or during implant, and post-operative persistent/symptomatic seroma and wound management.Patient risk factors may include diabetes, smoking, age, malnutrition, immunosuppressive therapy, post-operative instruction noncompliance, and hygiene.As with any surgical procedure, there are always risks of complications for surgical repair of hernias and soft tissue deficiencies, with or without mesh.These may include but are not limited to, adhesions and related harms, bleeding, bowel obstruction, compromised device biocompatibility, contamination which may lead to patient harms, device damage, dysphagia, erosion or extrusion and related harms, exposure or protrusion and related harms, fever, fistula, gerd recurrence, defect recurrence and related harms, ileus, increased procedure time and related harms, irritation or inflammation, infection, mesh migration, mesh contraction, pain, paresthesia, perforation, revision/re-intervention, seroma or hematoma and related harms, tissue ischemia, wound complications and wound dehiscence and additional intervention including surgery.Many of the potential complications are associated with the patient¿s underlying disease progression, co-morbidities, additional medical history and/or other surgical procedures.The above inherent risks are typically detailed in standard informed consent documents.The device was not able to be returned to gore for evaluation; therefore, a direct product analysis could not be conducted.Review of the manufacturing and sterilization records verified that the lot met all pre-release specifications.C1: the plus antimicrobial product coating contains silver carbonate [approximately 800 micrograms per cubic centimeter of product (g/cm3)], and chlorhexidine diacetate [approximately 1600 micrograms per cubic centimeter of product (g/cm3)].W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® DUALMESH® PLUS BIOMATERIAL
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY MOUNTAIN B/P
3750 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
jorja nackard
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key14881120
Report Number2017233-2022-03085
Device Sequence Number12302660
Product Code FTL
UDI-Device Identifier00733132601110
UDI-Public00733132601110
Combination Product (Y/N)Y
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K063435
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2002
Device Explanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/01/2004
Device Model Number1DLMCP04
Device Catalogue Number1DLMCP04
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/27/2022
Supplement Date Received by Manufacturer07/08/2022
07/08/2022
07/08/2022
Initial Report FDA Received Date06/30/2022
Supplement Report FDA Received Date07/11/2022
02/13/2023
02/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/01/2002
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age41 YR
Patient SexUnknown
Patient Weight89 KG
-
-