EDWARDS LIFESCIENCES LLC EDWARDS SAPIEN 3 ULTRA VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9750TFX23
Device Problem Coagulation in Device or Device Ingredient (1096)
Patient Problem Thrombosis/Thrombus (4440)
Event Date 09/29/2021
Event Type  Injury  
Event Description
Thv/tvt registry alternative summary reporting adverse event submission for events received by edwards during quarter 1 2022 for aortic serious injury events for the sapien 3 ultra valve.Thv/tvt registry alternative summary reporting for serious injury device thrombosis event associated with the sapien 3 ultra transcatheter heart valve, in the aortic position, for quarter 1 2022.This report is for device thrombosis serious injury event for sapien 3 ultra transcatheter heart valve.The age for this event is 74 years.The gender is as follows: female.
 
Manufacturer Narrative
Exemption number e2016006.This report provides data from thv/tvt registry data exemption number e2016006 and summarizes 1 of aortic serious injury device thrombosis for the sapien 3 ultra transcatheter heart valve in aortic position.The 'time to event' (tte, in days) for this event was 408.The device identification (di) numbers for edwards sapien 3 ultra transcatheter heart valve are: (b)(4).Per the instructions for use (ifu), valve thrombosis is a potential risk associated with the use of the transcatheter heart valve (thv).Valve thrombosis is a rare and well-recognized complication associated with the use of bioprosthetic heart valves.Valve thrombosis is the formation of significant blood clots forming on the valve.These clots could significantly impact the functionality of the valve resulting in heart failure or thromboembolism.Anticoagulation therapy must be evaluated patient by patient.The cause for valve thrombosis is thought to result from an interaction between components of blood and the prosthesis or turbulent blood flow in and around the prosthesis.Several factors can cause development of thrombosis, including: inadequate anticoagulant therapy after valve implant, atrial fibrillation, medications, cancerous tumors, systemic diseases (e.G., systemic lupus erythematosus, inflammation and damage to various body tissues, including joints, skin, kidneys, heart, lungs, blood vessels, and brain), and reduced cardiac pumping (low ejection fraction).Thrombosis has an extremely low incidence rate in bioprosthetic heart valves.Thrombosis is a complex process triggered by the interaction between the host and the device and is highly variable among patients.Since the mechanism of thrombotic deposition on bioprosthetic heart valves is not fully understood, the cause cannot always be confirmed.In this case, specific clinical details were not available, and it is not possible to determine potential contributing factors.A complaint history for this type of event is reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventive actions are required.
 
Manufacturer Narrative
Correction to the exemption number field per fda request on 07/07/2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 ULTRA VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key14881130
MDR Text Key295115261
Report Number2015691-2022-06476
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140031
Exemption Number2016006
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number9750TFX23
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age74 YR
Patient SexFemale
-
-