MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV SYS PED.W/SA 10 A.PRECHAMBER; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV SYS PED.W/SA 10 A.PRECHAMBER; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FV439T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 05/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that a progav shuntsystem (#fv439t) was implanted during a procedure performed on (b)(6) 2021.According to the complainant, the shunt system was believed to be difficult to adjust.The patient underwent a revision procedure on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 6 years.Gender: female.
 
Additional Manufacturer Narrative
Additional information - a4, b7, d4 (serial #), d9, e1 corrected information - b3, d6b investigation: visual inspection: the following observations were made during the visual inspection: -particles in the delivery liquid -deformation of the housing -deposits on and in the prechamber.Measurement of surface of the housing: to verify whether the deformation had an effect on the plane-parallelism of the housing surface, this was measured using a dial gauge.Deformation detected: yes - measured value: - 0.072 mm [tolerance (0 ± 0.02mm)] too much pressure on the valve, for example through the adjustment tool or resulting from a fall or hit to the head of the patient, can compromise the integrity of the valve.Permeability test: the test showed that the progav shunt system is permeable.Computer control test: the computer controlled test showed the opening pressure of the progav, at a reference flow rate of 20 ml/h in a horizontal position of the valve, to be 1.09 cmh2o.This is within the specified tolerance of 3 cmh2o ± 3 cmh2o.Additionally, the shuntassistant was tested according to standard procedure in the vertical position of the valve.At a fixed opening pressure of 10 cmh2o in the vertical position, a pressure of 10 cmh2o +4/-2 cmh2o is expected.The results indicated that at a reference flow of 20 ml/h in the vertical position, the shuntassistant had a pressure of 10.73 cmh2o.This is within the specified tolerance of 10 cmh2o +4/-2 cmh2o.Adjustability test: the progav was found to be adjustable to all pressure settings.The shuntassistant is a fixed pressure valve, therefore the adjustability test is inapplicable.Braking force and brake function test: the braking force and brake function test investigates whether the brake function of the adjustable valves is present and how much force must be exerted on the housing to release the rotor to adjust the valves using the integrated magnet of a specific measurement apparatus of braking force.The braking force of the progav was not within the specified tolerance.The brake could be deactivated with a lower force on the valve.The brake function is present.Internal inspection of product.In order to verify whether the investigated shunt system were compromised by the known risks of hydrocephalus therapy, e.G.By a build-up of natural substances (protein, blood, or tissue particles) in the cerebrospinal fluid, the valves are finally opened with a special tool.After dismantling of the valves, deposits were found in both valves.Results: in advance, we are obliged to inform you that the condition of the product could be affected due to the interval between the explantation date and therelease of the product.So the investigation result could be invalid because the time of delay of the start of the investigation.Based on our investigation results, we have determined that there was a deformation oft he progav valve housing and a lower braking force at the progav valve at the time of our investigation.The detected deformation was the cause of the brake malfunction.Additionally, the existing deposits in the valves can temporarily impair the function and is described as concomitant symptoms in hydrocephalus therapy.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.From our point of view, no further regulatory actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV SYS PED.W/SA 10 A.PRECHAMBER
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key14881134
Report Number3004721439-2022-00201
Device Sequence Number12302665
Product Code JXG
UDI-Device Identifier04041906132521
UDI-Public4041906132521
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K103003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/22/2024
Device Model NumberFV439T
Device Catalogue NumberFV439T
Device Lot Number20043983
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2022
Supplement Date Received by Manufacturer11/09/2022
Initial Report FDA Received Date06/30/2022
Supplement Report FDA Received Date12/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age6 YR
Patient SexFemale
Patient Weight48 KG
-
-