CHRISTOPH MIETHKE GMBH & CO KG PROGAV VALVE; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FV410T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 05/31/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a progav (#fv410t) was implanted during a procedure performed on (b)(6) 2013.According to the complainant, the shunt system was believed to be operated in overdrainage and had adjustment difficulties.The patient underwent a revision procedure performed on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 9 years.Height: 140 centimeters (cm).Weight: 45 kilograms (kg).Gender: female.
 
Additional Manufacturer Narrative
Investigation: visual inspection: a deformation of the outer housing of the progav valve was observed through the visual inspection.The measurement of the plane parallelism could confirm that with a value of -0,054mm the housing membrane is clearly outside tolerance (0 ± 0.02 mm).Permeability test: a permeability test has shown that the valve is permeable.Klick force and brake function test: the brake functionality test has shown that the brake function did not operate as expected.The results indicate that the rotor no longer performs as required.Computer controlled test: to investigate the claim of over/under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.Due to the brake failure, the valve can´t hold the adjusted opening pressure.Therefore, the computer-controlled test is inapplicable.Internal inspection: after dismantling the valve, deposits were found in progav.Results: based on our investigation results, we can determine a defective brake as well as deposits in the valve.The observed deformation of the housing can be named as the cause of the defective brake.In addition, the visible deposits in the valve can lead to a functional impairment.Deposits caused by substances naturally present in the body, such as protein, blood or tissue particles, are one of the known and inevitable risks and side effects of hydrocephalus therapy.Even small amounts of protein can compromise the integrity of the valve.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV VALVE
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key14881187
Report Number3004721439-2022-00181
Device Sequence Number20409914
Product Code JXG
UDI-Device Identifier04041906132071
UDI-Public4041906132071
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K103003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFV410T
Device Catalogue NumberFV410T
Device Lot Number20020061
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2022
Initial Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/21/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age9 YR
Patient SexFemale
Patient Weight45 KG
-
-