MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 168 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 168 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Model Number 168-650
Medical Device Problem Codes Fluid/Blood Leak (1250); Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Dhr trend summary: no patterns were found; therefore, no additional actions are deemed required.The trend of a050401 - fluid leak complaints will continue to be monitored and corrective actions will be taken in the future if deemed appropriate.The event of "capsular contracture, baker grade iv" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation and capsular contracture, baker grade iv.
 
Event or Problem Description
Patient reported left side deflation and exchange from textured to smooth implants due to the patients concerns with the product.Additionally, healthcare professional reported left side "capsular contracture baker grade iv." device remains implanted.
 
Additional Manufacturer Narrative
Device evaluation:analysis of the returned device identified: crease flat, delamination of the plug strap disk shell and opening curved on posterior.Leak test and microscopic analysis was performed which identified: sharp opening on posterior, delamination of the plug strap disk shell and non-penetrating nick.The fill test inspection was performed, the result is no blockage.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a sharp opening on posterior assessed as an unidentified (tear) opening.A delamination total of the plug strap disk shell (adhesive failure).
 
Event or Problem Description
Healthcare professional reported left side hole non specific-in inferior aspect.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 168 SALINE FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key14881451
Report Number9617229-2022-10807
Device Sequence Number6307094
Product Code FWM
UDI-Device Identifier10888628000919
UDI-Public10888628000919
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P990074
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/15/2019
Device Model Number168-650
Device Catalogue Number168-650
Device Lot Number2843498
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/06/2022
Supplement Date Received by Manufacturer07/08/2022
Initial Report FDA Received Date06/30/2022
Supplement Report FDA Received Date08/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/15/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight73 KG
-
-