| Model Number |
DFR00V |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Discomfort (2330); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a female patient was implanted with a synergy intraocular lens (iol) three months ago.The quality of the patient's vision has been poor since the surgery, and worse than pre-operative (op).Surgeon can get her to 20/25 very slowly, but worse than pre-op and worse than left eye (os) with cataract.Surgeon cannot find anything wrong with the eye, and plans on doing a lens exchange.Cornea: patient had lasik, but cornea was clear with normal mires, and spherical.The implant is well centered, optical coherence tomography (oct) is normal macula, the capsule is mild opacity which is not significant, and the surgeon does not want to do a yttrium aluminum garnet (yag) if they are going to exchange the lens.Through follow-up, it was learned the iol was explanted from the patient's right eye.Another johnson & johnson lens, model dib00 19.0 diopter was implanted as a replacement.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Information unknown/not provided.Date of event: exact date unknown/not provided.Best estimate date is 3-4 months ago (february - march 2022).Brand name: unknown only provided as synergy, the serial number was not provided.Model number: unknown, as the serial number was not provided.Catalog number: unknown, as the serial number was not provided.Serial number: unknown, as information was not provided.Expiration date: unknown, as the serial number was not provided.Udi number: unknown, as the serial number was not provided.If implanted; give date: unknown/not provided.Best estimate date is 3-4 months ago (february - march 2022).Establishment name: unknown/not provided.Address, city, state, zip code: unknown/not provided.Telephone number: unknown/not provided.The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Device manufacture date: unknown, as the serial number was not provided.(b)(4).Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional info: through follow-up the account information and product information has been provided.Therefore, the following fields that were previously reported as unknown have now been updated accordingly.The following section has been updated accordingly: section d1: brand name: (b)(6) section d4: model number: dfr00v.Section d4: catalog number: dfr00vu190.Section d4: serial number: (b)(6).Section d4: expiration date: aug 11, 2024.Section d4: udi number: (b)(4).Section e1: establishment name: (b)(6).Section h4: device manufacture date: aug 11, 2021.Corrected data: through follow-up, the product identifiers were provided, therefore, this supplemental filing is to correct sections d2.The following section has been updated accordingly: section d2: common device name: intraocular lens section d2: device product code: hql.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|
|