MAUDE Adverse Event Report: SMITH & NEPHEW, INC. CUFF STITCH 180 DEG RIGHT; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. CUFF STITCH 180 DEG RIGHT; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 7209234
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy lab/demo procedure, the cuff stitch broke into pieces when passing the tendon.It broke inside the patient but all the broken pieces were retrieved.The procedure was successfully completed without surgical delay using a back-up device.No patient complications were reported.The current health status of the patient is stable.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy procedure, the cuff stitch broke into pieces when passing the tendon.It broke inside the patient but all the broken pieces were retrieved.The procedure was successfully completed without surgical delay using a back-up device.No patient complications were reported.The current health status of the patient is stable.
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A complaint history review found no similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A visual inspection of the returned device found that it is not in its original packaging.The entire distal tip is fractured away.The handle shows wear from use with the lot number clearly visible.The impact to the patient beyond that which has already been reported cannot be confirmed nor concluded.Therefore, no further medical assessment is warranted at this time.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).H10 h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no way to determine if the device contributed to the reported event.A complaint history review found no similar reported events.Insufficient product identification information was provided, and thus, a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.This case reports, the cuff stitch broke inside the patient when passed through the tendon.All pieces were retrieved, and a smith and nephew backup device was used to successfully complete the procedure with no damage caused and no procedural delay.The method of the device retrieval was not provided.The patient¿s current health status is stable.Without the requested supporting clinical information or the device, a thorough medical investigation cannot be rendered.The impact to the patient beyond that which has already been reported cannot be confirmed nor concluded.Therefore, no further medical assessment is warranted at this time.The complaint was not confirmed.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The entire distal tip is fractured away.The handle shows wear from use with the lot number clearly visible.There was a relationship found between the device and the reported event.A complaint history review found no similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
 
Event or Problem Description
It was reported that during an arthroscopy procedure, two cuff stitch broke into pieces when passing the tendon.They broke inside the patient but all the broken pieces were retrieved.The procedure was successfully completed without surgical delay using a back-up device.No patient complications were reported.The current health status of the patient is stable.
 
Additional Manufacturer Narrative
H2: corrected data on h6 (health effect - impact code).
 
Event or Problem Description
It was reported that during an arthroscopy procedure, two cuff stitch broke into pieces when passing the tendon.They broke inside the patient but all the broken pieces were retrieved.The procedure was successfully completed without surgical delay using a back-up device.No patient complications were reported.The current health status of the patient is stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUFF STITCH 180 DEG RIGHT
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14881946
Report Number1219602-2022-00932
Device Sequence Number6245182
Product Code NBH
UDI-Device Identifier03596010473127
UDI-Public03596010473127
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7209234
Device Catalogue Number7209160
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2022
Supplement Date Received by Manufacturer07/06/2022
07/06/2022
04/19/2023
10/25/2023
01/08/2024
Initial Report FDA Received Date06/30/2022
Supplement Report FDA Received Date07/07/2022
07/07/2022
04/25/2023
10/25/2023
01/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age24 YR
Patient SexMale
-
-