• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS AMBICOR PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS AMBICOR PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number 72401450
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an ambicor penile prosthesis (app) surgical procedure due to a fluid leak caused by a hole in the cylinder.There were no patient complications reported.
 
Event or Problem Description
It was reported that the patient underwent an ambicor penile prosthesis (app) surgical procedure due to a fluid leak caused by a hole in the cylinder.There were no patient complications reported.
 
Additional Manufacturer Narrative
Upon receipt at our quality assurance laboratory, this ambicor penile prosthesis (app) underwent a thorough analysis.The pump and cylinders were visually and microscopically examined.No anomalies were found with the pump or cylinders.Based on the information available and analysis results, product investigation was unable to confirm the reported clinical observations of hole/perforated and fluid leak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS AMBICOR PENILE PROSTHESIS
Common Device Name
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key14882511
Report Number2124215-2022-23732
Device Sequence Number5089453
Product Code FHW
UDI-Device Identifier00878953000947
UDI-Public00878953000947
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/02/2026
Device Model Number72401450
Device Catalogue Number72401450
Device Lot Number1000520050
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2022
Supplement Date Received by Manufacturer07/14/2022
Initial Report FDA Received Date07/01/2022
Supplement Report FDA Received Date07/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-