We checked the returned unit and confirmed that the nozzle gluing missing.Based on the result, we concluded that it was caused due to the physical damage applied on the nozzle gluing.In addition, we confirmed that the insertion flexible tube (ift) coating damage, the operation channel leak, the bending rubber cut, the operation channel kink, the suction channel kink, the angle wire grinding, the insertion flexible tube (ift) dented, and the air/water socket chipped/cut o-ring ; however, they are not the main cause, and/or irrelevant to the alleged complaint.As a result of confirming the detail as gfe, no response was obtained, so we can not deny the possibility of the glue dropped in human body.Therefore, based on the technical report ""hr-rpt-0585(nozzle)"" and/or the risk analysis results, it was evaluated to submit mdr.
|