As initially reported to customer relations: a female patient in her 60s with compression of the common iliac and external iliac veins on right side underwent an ivus and venography procedure in which the zilver vena venous self-expanding stent, g57449 and g57448, were used.Access site was right groin ivus and venography was performed physician selected 16x140 vena to stent from the confluence into the external iliac vein on right side.No predilation was done.After deploying stent optimally, physician advanced a bard atlas 16x60 balloon into the stent.The tech who was scrubbed in states the physician was not stepping on fluoro while advancing balloon, and once he did step on fluoro both the physician and tech saw that the stent had almost completely, about 95%, migrated into the patient¿s ivc (pr365969).They both agreed that the balloon catheter bumped the caudal edge of the stent and caused it to migrate significantly.Physician attempted to use the bard atlas balloon to pull the stent back down into the common iliac, but was unsuccessful.Then, physician deployed another 16x140 vena into common iliac, overlapping the 2 stents a few centimeters.Due to stent migration, a third stent, the 16x100, was needed to extend into the external iliac.Physician used same 16x60 atlas balloon to post dilate the stents, and as he was advancing the balloon into the 16x100 stent, the balloon bumped into this stent also and caused it to jump about 2 cm but was able to use the balloon to pull the stent back down to original landing zone (pr366108).Did any unintended section of the device remain inside the patient¿s body? if yes, please describe.Was the patient hospitalized or was there prolonged hospitalization? did the patient require any additional procedures due to this occurrence? if yes, please describe.Did the product cause or contribute to the need for additional procedures? if yes, please specify additional procedures and provide details.Has the complainant reported any adverse effects on the patient due to this occurrence? has the complainant reported that the product caused or contributed to the adverse effects? please specify adverse effects and provide details.District managers reply on 16june2022: th 17june2022.Did any unintended section of the device remain inside the patient¿s body? if yes, please describe.No.Was the patient hospitalized or was there prolonged hospitalization? no.Did the patient require any additional procedures due to this occurrence? if yes, please describe.No.Did the product cause or contribute to the need for additional procedures? no.If yes, please specify additional procedures and provide details.Has the complainant reported any adverse effects on the patient due to this occurrence? no.Has the complainant reported that the product caused or contributed to the adverse effects? please specify adverse effects and provide details.No.Prefix zvt7.Are images of the device or procedure available? n/a, yes, no yes will send --- see attached.Did the patient have pre-existing conditions? n/a, yes, no.If yes, please specify: yes history of multiple dvts.Please describe the native state of the vessel (i.E.Was the anatomy tortuous? straight vessel.Was the vessel fibrotic?) n/a, tortuous, calcified, fibrotic, other if other, please specify: somewhat fibrotic.Was a stent previously placed during previous procedures? n/a, yes, no no, just 3 in this procedure.Was the device used percutaneously? n/a, yes, no yes.Where on the patient was the percutaneous access site? right groin.Was the access site jugular or femoral? n/a, jugular, femoral other.If other, please specify: fem.What disease pathology was being treated? may thurner, acute or chronic obstruction, restenosis, other? compression./may thurner.If other, please specify.Was the lesion approached via contralateral or ipsilateral? n/a, contralateral, ipsilateral ipsilateral.Was pre-dilation performed ahead of placement of the stent? n/a, yes, no no.What was the target location for the stent? right external iliac into common fem.Details of access sheath used (name, fr size, length)? terumo, unsure of size was the device flushed through both flushing ports before the procedure, as per ifu? n/a, yes, no yes.Details of the wire guide used (name, diameter, hyrdophyllic)? standard 035 wire, unsure of kind.Was resistance encountered when advancing the wire guide to the target location? n/a, yes, no no.Was resistance encountered when advancing the delivery system to the target location? n/a, yes, no no.If resistance was met, how did the physician address this? did the tip of the delivery system cross the target location? n/a, yes, no yes.Did the user pull the handle towards the hub during deployment, per ifu? n/a, yes, no yes.Did the user push the hub during deployment? n/a, yes, no.Did the user remove slack in the delivery system before deployment, per ifu? n/a, yes, no yes was the stent deployed smoothly / without resistance? n/a, yes, no yes.Was the stent fully deployed in the patient? n/a, yes, no yes.Was the stent fully deployed before removing the delivery system from the patient? n/a, yes, no yes.Was post dilation performed after the placement of the stent? n/a, yes, no yes.Was the delivery system damaged/kinked/twisted during deployment? n/a, yes, no no.What intervention (if any) was required? upon advancing the bard atlas balloon into this stent, the actual balloon caused this stent to move up a couple centimeters up.Balloon was then inflated and used to bring stent back down.Was the secondary intervention performed during the same procedure as the device failure or was it scheduled for another day? n/a, same procedure, another day same.Were any other defects (other than the complaint issue) observed on the delivery system prior to return (e.G.Kink)? n/a, yes, no.Please specify if yes.No.
|