Report confirmed.Evaluation determined that leg of the device was bent.The likely cause of failure is inadequate preventative maintenance.It was also noted the bearings, retaining ring and o-ring are worn.Also, the color band is missing and the laser markings are partially faded.The user manual contains the following warnings ¿do not use a legend attachment if any part of the attachment appears to be bent, loose, missing or damaged.Do not use excessive pressure, such as bending or prying, on attachments or dissecting tools.This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff." in addition, ¿insert dissecting tool into motor collet with a slight rotational motion.A tactile and audible click is observed indicating that the dissecting tool is fully seated.The maximum specified service interval is 24 months.Device has been in use for approximately 77 months with no record of factory service during this period.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
|