It was reported that an erbe system, argon plasma coagulator (apc)/electrosurgical unit (esu/generator, model vio 3, part number (p/n) 10160-000 (b)(4)) system was involved in a patient incident during an upper gastrointestinal procedure.The patient was being treated with argon plasma coagulation to control/stop bleeding in the patient's stomach.The apc/esu system was used with an fiapc straight fire, 2.3 mm probe (part number 20132-214).The equipment settings were pulsedapc mode, effect 2.5, 0.8 liter/minute flow rate.A perforation occurred and surgical intervention was required to address the issue.
|
There is a plan to have the erbe equipment checked/evaluated at erbe usa.If any issues are found, a follow up report will be provided.No anomalies were found in device history records (dhrs) of the involved devices.Most likely, there were many factors involved in the event.At this time, no determination could be made as to the cause of the incident.
|