| Model Number |
24690 |
| Medical Device Problem Codes |
Material Puncture/Hole (1504); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft perforation occurred.A 7.0mmx30mmx80cm (4f) sterling balloon catheter was prepared for use.However, prior to advancing over the guidewire, the physician noticed that the shaft of the balloon was torn.Additionally, it was noted that the device was kinked.The procedure was completed with a 7x20 sterling balloon catheter.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon, and tip were visually and microscopically examined.A kink was visible 55.8 cm from the hub.The balloon was inflated to 14atm and held for one minute.There was no further damage to the sterling balloon.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft perforation occurred.A 7.0mmx30mmx80cm (4f) sterling balloon catheter was prepared for use.However, prior to advancing over the guidewire, the physician noticed that the shaft of the balloon was torn.Additionally, it was noted that the device was kinked.The procedure was completed with a 7x20 sterling balloon catheter.No patient complications were reported.
|
| |
|