| Model Number |
08P1934 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed a falsely elevated magnesium result on the alinity c processing module, serial number (b)(4).Repeat testing was lower on a different alinity c processing module.Sid (b)(6): initial result: 0.73 mmol/l, repeat: 0.49 mmol/l.Normal range: 0.66 to 1.07 mmol/l.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
All available patient information was included.Additional patient details are not available.Trending review determined no adverse trend for the issue for the product.Historical complaint review determined there is normal complaint activity for the lot number.Device history record review did not show any potential non-conformances, or deviations.Labeling was reviewed and found to adequately address the issue under review.Worldwide data from abbottlink was reviewed and determined that patient median result for the lot is comparable with other lots in the field and within established baselines confirming no systemic issue for the lot.Based on the investigation, no deficiency for lot number 33108ud00 was identified.
|
| |
|