MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CRS RP INSRT SZ 5 8MM; KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CRS RP INSRT SZ 5 8MM; KNEE TIBIAL INSERT Back to Search Results
Model Number 1517-10-508
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Discomfort (2330)
Date of Event 06/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision completed secondary to thigh pain.No evidence of infection or loosening.Surgeon suspected the pain may be being generated from a large femoral press fit stem.After discussions with the patient, the decision was made to do a femoral only revision - plan a being to revise the femoral sleeve to a larger size and reduce the femoral stem diameter.Plan b would be either an s-rom hinge or lps distal femoral replacement depending on bone quality following removal of the current attune revision femoral components.The removal process was not simple as the initial components were well fixed, and as a result there was a complete fracture of the lateral distal femoral condyle as well as significant metaphyseal bone loss of the distal femur whilst attempting to remove the well fixed sleeve and stem.Palacos bone cement was used in all procedures.Plan b was actioned and a lps distal femoral replacement was performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.All available photo evidence were reviewed and does not shows an implant fracture, disassociation or any indicative related to a device nonconformance.Based on the information provided, a defect or malfunction cannot be established, therefore the allegation is not confirmed.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Event or Problem Description
Additional information received indicated that there was no surgical delay reported and the affected side involved in this event was the right knee.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CRS RP INSRT SZ 5 8MM
Common Device Name
KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14886483
Report Number1818910-2022-12216
Device Sequence Number6283438
Product Code NJL
UDI-Device Identifier10603295055914
UDI-Public10603295055914
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1517-10-508
Device Catalogue Number151710508
Device Lot Number9058832
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2022
Supplement Date Received by Manufacturer09/13/2022
Initial Report FDA Received Date07/01/2022
Supplement Report FDA Received Date09/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/25/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CRS FEMORAL RT SZ 5 CEM; ATUN FEM SLV M/L 45MM FULL POR; ATUNE PRESSFIT STR STEM18X60MM; PALACOS BONE CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-