• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ULTRATHANE COPE NEPHROURETEROSTOMY SET; FAD STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ULTRATHANE COPE NEPHROURETEROSTOMY SET; FAD STENT, URETERAL Back to Search Results
Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Occupation: lab manager.Pma/510(k) #: k171603.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported a patient required a ultrathane cope nephroureterectomy set for a bilateral drain exchange in the kidneys.During the procedure, the stiffener's hub separated upon removal.The device was replaced with another device of the same lot, and the suture broke upon locking the pigtail.The device was removed, and the procedure was completed by placing two similar devices for bilateral drainage.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged or unavailable.Correction: d9, h3 investigation ¿ evaluation.On 13jun2022, cook medical received a complaint from a representative at the (b)(6) medical center located in the city of (b)(6).Upon the replacement of a previous ultrathane cope nephroureterostomy catheter with a new ultrathane cope nephroureterostomy set (rpn: ult8.5-8.5-24-nucl-b-rh, lot: 13968635) over a wire guide, the hub to the supplied flexible stiffener separated upon removal from the catheter.The stiffener was outside the patient, and the catheter was subsequently removed from the patient.The patient did not experience any adverse effects due to this occurrence.Reviews of the documentation, including the complaint history, device history record, drawing, instructions for use (ifu) and quality control procedures of the device, were conducted during the investigation.The complaint device was not returned for evaluation; therefore, no physical examinations could be conducted.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that the sufficient controls are in place to detect this failure prior to release.A review of the device history record (dhr) for lot 13968635, catheter subassembly lots (sa13816202 and sa13816205), radiopaque band tubing lots (sa13816239 and sa13878953) and raw material lots found no nonconformances that could have contributed to the reported failure mode.Nonconformances relevant to the failure were found for subassembly lot sa13816239 for ¿shaft difficult to wire" (qty 5) and sa13878953 for ¿tight id¿ (qty 1 ), in which all nonconforming devices were scrapped prior to further processing.It should be noted that there was one other complaint associated with the final product lot number regarding suture breakage.Cook also reviewed product labeling.Per the ifu, t_nucl_rev4, under the heading precautions: when inserting a stiffening cannula into the catheter with retention suture, hold suture during cannula insertion to avoid bunching or tangling of the suture.Evidence provided upon review of the dmr, ifu and dhr suggests that the device was not manufactured out of specification, and that there are no nonconforming devices in house or out in the field.Based on the information provided, no returned product and the results of our investigation, it was concluded that a component failure without any design or manufacturing issue contributed to the reported event.It is feasible to suggest that the suture string became tangled around the stiffener due to the string being improperly held.As the complaint device was not returned, this cannot be confirmed.The appropriate personnel have been notified.Per the risk assessment no further action is required.We will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding patient and/or event details has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE COPE NEPHROURETEROSTOMY SET
Common Device Name
FAD STENT, URETERAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14886897
Report Number1820334-2022-01146
Device Sequence Number17691330
Product Code FAD
UDI-Device Identifier00827002481763
UDI-Public(01)00827002481763(17)240518(10)13968635
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/18/2024
Device Model NumberN/A
Device Catalogue NumberULT8.5-8.5-24-NUCL-B-RH
Device Lot Number13968635
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2022
Supplement Date Received by Manufacturer12/15/2022
Initial Report FDA Received Date07/01/2022
Supplement Report FDA Received Date01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-