• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: QUIDEL CORPORATION LYRA DIRECT SARS-COV-2 ASSAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

QUIDEL CORPORATION LYRA DIRECT SARS-COV-2 ASSAY Back to Search Results
Model Number Coronavirus
Medical Device Problem Codes False Positive Result (1227); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation conclusion: returned product was tested in-house and worked as expected.No product performance issues were found.Customer data was reviewed, and analysis did not conclude a product performance issue.Analysis revealed a higher fam signal attributed to post-run analysis of the data.Quidel provided modified programming for the third-party equipment to mitigate the issue moving forward.Root cause: third- party equipment programming, source: email.
 
Event or Problem Description
One false positive sars result.The customer ruled out environmental contamination sources.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LYRA DIRECT SARS-COV-2 ASSAY
Common Device Name
LYRA DIRECT SARS-COV-2 ASSAY
Manufacturer (Section D)
QUIDEL CORPORATION
2005 east state street
athens OH 45701
Manufacturer Contact
karl luke
2005 east state street
athens, OH 45701
7405893382
MDR Report Key14887165
Report Number0002024674-2022-09063
Device Sequence Number5162878
Product Code QJR
UDI-Device Identifier30014613339144
UDI-Public30014613339144
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EUA200423
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCoronavirus
Device Catalogue NumberM124
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/28/2020
Initial Report FDA Received Date07/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-