| Model Number |
Coronavirus |
| Medical Device Problem Codes |
False Positive Result (1227); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
One false positive sars result.The customer ruled out environmental contamination sources.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation conclusion: returned product was tested in-house and worked as expected.No product performance issues were found.Customer data was reviewed, and analysis did not conclude a product performance issue.Analysis revealed a higher fam signal attributed to post-run analysis of the data.Quidel provided modified programming for the third-party equipment to mitigate the issue moving forward.Root cause: third- party equipment programming, source: email.
|
| |
|
Search Alerts/Recalls
|