• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012276-15
Medical Device Problem Codes Material Rupture (1546); Improper or Incorrect Procedure or Method (2017); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a 99% stenosed, heavily calcified and moderately tortuous vessel in the proximal left anterior descending (plad) artery.The 3.5x15mm trek rx balloon dilatation catheter (bdc) had resistance due to the heavily calcified part at the entrance of the proximal lad but crossed the lesion.During inflation at 4 atmospheres (atms), the balloon was not pressurized and deflated.There was no resistance during withdrawal.The device was not soaked prior to use and the device was not prepared (air aspiration) outside the anatomy prior to use.There was no difficulty removing the stylet or protective sheath.A non-abbott balloon was used to complete the procedure.There was no adverse patient effect.There was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
(b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.It was reported that the device was prepped inside the anatomy.It should be noted that the coronary dilatation catheters (cdc), trek rx and mini trek rx, global instructions for use (ifu) specifies: all air must be removed from the balloon and displaced with contrast prior to inserting into the body, otherwise, complications may occur.It is unknown if the ifu violation caused or contributed to the reported complaint.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14887560
Report Number2024168-2022-07161
Device Sequence Number6282901
Product Code LOX
UDI-Device Identifier08717648138577
UDI-Public08717648138577
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model Number1012276-15
Device Catalogue Number1012276-15
Device Lot Number10819G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/14/2022
Initial Report FDA Received Date07/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-