• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIGC60VM
Medical Device Problem Codes Failure to Form Staple (2579); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Unspecified Tissue Injury (4559)
Date of Event 06/21/2022
Type of Reportable Event Serious Injury
Event or Problem Description
According to the reporter, on a laparoscopic sleeve gastrectomy, during stomach transection, the device fired but the entire staple line of five reloads were bleeding.There was poor staple formation.The entire staple line was clipped to stop the bleeding.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
D10 concomitant product: sigc60vm, tri 2.0 sigc60vm 60 vsclr med cartridge (lot#n1c0014y); sigc60vm, tri 2.0 sigc60vm 60 vsclr med cartridge (lot#n1c0014y); sigc60vm, tri 2.0 sigc60vm 60 vsclr med cartridge (lot#n1c0014y); sigc60vm, tri 2.0 sigc60vm 60 vsclr med cartridge (lot#n1c0014y); siglu60a, tri 2.0 siglu60a 60 load unit (serial#unknown); unk-egiaadapter, unknown egia adapter (serial#unknown) additional information: d9, g1(manufacturer name, mfr contact first name, last name, street 2, mfr city, region, postal code, email and phone number), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted the staple cartridge was fully fired with the sled fully advanced.Functionally, the cartridge was loaded into a representative handle, adapter, and loading unit.The cartridge and representative loading unit were loaded onto a representative manual handle.The interlock was overridden and the cartridge was applied to test media.Test media was cleanly transected.The interlock was tested and found to function properly.It was reported that the staples did not form as expected and there was unexpected bleeding occurred along the staple line.The reported issues were confirmed.The most likely cause was not traced to the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
5920 longbow drive
8200 coral sea st ne
mounds view,mn, CO 55112
7635265677
MDR Report Key14887864
Report Number1219930-2022-02552
Device Sequence Number20121418
Product Code GDW
UDI-Device Identifier10884521543560
UDI-Public10884521543560
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K160176
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model NumberSIGC60VM
Device Catalogue NumberSIGC60VM
Device Lot NumberN1C0014Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/21/2022
Supplement Date Received by Manufacturer10/07/2022
Initial Report FDA Received Date07/01/2022
Supplement Report FDA Received Date10/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PLEASE SEE NOTE ON H10
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-