HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP/DRIVELINE; VENTRICULAR (ASSISST) BYPASS
|
Back to Search Results |
|
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Stroke/CVA (1770); Hemorrhage/Bleeding (1888); Intracranial Hemorrhage (1891); Unspecified Infection (1930); Ischemia Stroke (4418); Thrombosis/Thrombus (4440); Unspecified Heart Problem (4454)
|
Event Date 02/19/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
This information is based entirely on journal literature.Medtronic was made aware of this event through a search of literature publications.This event occurred outside the us.Patient information is limited due to confidentiality concerns.Of note, multiple patients and multiple manufacturers were noted in the article; however, a one-to-one correlation could not be made with unique product serial/lot numbers.The baseline gender/age characteristics is male/55 years old.The model listed in the report is a representative of the model family, as there is no specific model listed.Without a lot number or device serial number, the manufacturing date cannot be determined.Since no device id was provided, it is unknown if this event has been previously reported.A request for additional information will be made and upon receipt a supplemental report will be submitted accordingly.Referenced article: impact of frailty on mortality and morbidity in bridge to transplant recipients of contemporary durable mechanical circulatory support.The journal of heart and lung transplantation, vol 41, no 6, june 2022.Doi: 10.1016/j.Healun.2022.02.008 related to report number: 3007042319-2022-06374 investigation of this event is pending, and a supplemental report will be sent upon its completion.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
A journal article was reviewed that contained information regarding the impact of frailty on mortality and morbidity in bridge to transplant patients with mechanical circulatory support.The article reports patients who underwent ventricular assist device (vad) or biventricular (bivad) implants and experienced adverse events such as gastrointestinal bleeds (gib), intracranial hemorrhages, and other bleeds, ischemic and embolic strokes, pump thrombus, driveline infections, systemic bacteremia, worsening right ventricular (rv) function, and other unknown complications.Patients were categorized into frail and non-frail groups with some patients had longer intensive care unit (icu) and hospital stays or required longer ventilator time.Patients were medically managed, required repeated cardioversion, or other types of intervention.The status/disposition of the vads is unknown.No further patient complications have been reported as a result of this event.Further follow up did not yet yield any additional information.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for investigation completion.Product event summary: one (1) pump with unknown serial number was not returned for evaluation.Log file analysis could not be performed since log files covering the reported event date were not available for analysis.Review of the sterility certificate could not be conducted since the serial number is unknown.Based on the available information, the device may have caused or contributed to the reported event.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patient's complex post-operative course, including care of the driveline exit site.There are possible patient, pharmacological and procedural factors that may have contributed to this event.Investigation of this event is completed, and the file will be closed.If new information is received, the file will be re-opened, and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|