The complaint device was not to be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.The production site checked the batch record of impacted safety needle 25g x 1 ½", ref 900-7656, lot 02103067 with no anomalies found.No samples were received by the customer for evaluation.Archive sample evaluation: traction test: 10pcs have been checked for traction test between cannula and needle hub.All samples are within product specification requirements (max.Deflection 0.4mm for tw).Resistance to breaking test: 10pcs have been tested in accordance with iso 9626:2016 annex c.All samples passed the test.Stiffness test: 10pcs have been tested in accordance with iso 9626:2016 annex b.All samples are within product specification requirements (max.Deflection no anomalies were found on archive samples checked.Risk statement: the overall risk assessment, given the severity assessment of s3-moderate injury and a probability assessment of p1-very low, the final risk priority number is 3.Risk level is in green area/low risk according to our risk management procedure.
|