|
(mfr contact first name, last name, street1, mfr city, region, postal code, email, phone number) evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted the device was found to be damaged as a result of rough handling.It was reported that the obturator was damaged and the physical appearance of the product was different than expected.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the device has excessive force applied during clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: make an incision large enough to accommodate the size of the trocar sleeve at the site of the placement.Bluntly dissect through the fascia and peritoneum in a routine approach for open laparoscopy.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
According to the reporter, during a laparoscopic colectomy procedure, after opening the package and before using the product for the patient, the surgeon noticed that there was a dent with the obturator.The surgeon stopped using the device that was replaced after to complete the case and resolve the issue.There was no patient injury.
|