SMITH & NEPHEW, INC. JRNY II CR ISRT TRL RT SZ 7-8 9M; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Model Number |
74033671 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during tka, during trialing polys, the plastic of the bottom where the tracks are engaged into the jrny ii cr isrt trl rt sz 7-8 9m chipped.Surgeon was unable to get this poly seated after trying numerous times.After opening up another tray with same size poly, surgeon could seat the poly in the first try.No further complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that during tka, during trialing polys, the plastic of the bottom where the tracks are engaged into the jrny ii cr isrt trl rt sz 7-8 9m chipped.Surgeon was unable to get this poly seated after trying numerous times.After opening up another tray with same size poly, surgeon could seat the poly in the first try.No further complications or significant delay were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirms several pieces of the device are chipped off, rendering the device inoperable.The pieces were not returned.The device shows signs of significant wear and use.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it can no longer fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|