• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. KIT MMT-1760KPK 670G PATHWAY BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. KIT MMT-1760KPK 670G PATHWAY BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL Back to Search Results
Model Number MMT-1760KPK
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is part of a retrospective review and remediation efforts in response to a warning letter.Correction/removal number: z-0956-2020.(b)(4).S/w version 4.10e.The pump passed the displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test, self test and displacement accuracy test.No unexpected pump error 54 or pump error 3 alarms noted during testing however, the formatted history file confirmed pump error 54 (line number 1421 file number 32122) and pump error 3 alarms due to a software error.The pump was connected to a test transmitter/sensor and monitored without any connection interruptions.Pump and test transmitter/sensor calibrated successfully.Pump was connected to a test meter, contour next link, and was able to connect.A controlled glucose test solution was tested using the meter and sent to the pump without any issues.No sg vs bg anomalies noted during testing.Displacement test: 0.03475, pass.Rewind test: pass.Prime/seating test: pass.Basic occlusion test: pass.Force sensor test: 2.75, pass.Occlusion test: 1.700, pass.Self test: pass displacement accuracy test: 0.08685, pass.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
Medtronic received information that pump error 3, pump error 54 occurred.There was no adverse impact or consequence reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT MMT-1760KPK 670G PATHWAY BLACK MG
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
7635140379
MDR Report Key14898364
Report Number2032227-2022-260352
Device Sequence Number17651575
Product Code OZP
UDI-Device Identifier00763000096984
UDI-Public(01)00763000096984
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 07/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-1760KPK
Device Catalogue NumberMMT-1760KPK
Device Lot NumberHG2E0C3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/04/2018
Initial Report FDA Received Date07/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/15/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0956-2020
Patient Sequence Number1
Patient SexUnknown
-
-