• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON® SPINAL SYSTEM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON® SPINAL SYSTEM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 6642002
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding an event which occurred during a l3/4/5 olif procedure in a patient diagnosed with l3/4/5 vertebral canal stenosis.It was reported that the set screw was attached to the set screw retaining driver and final tightening was performed.Piece of the set screw threading was stuck to the set screw retaining driver and could not be removed.There was no patient symptom reported.
 
Additional Manufacturer Narrative
Product analysis: part # 6642002, lot # em16k037 visual and optical inspection confirmed the hex tip of the retaining driver has been stripped/damaged.The inner obturator has also been bent.This type of failure is consistent with torsional overload.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON® SPINAL SYSTEM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key14904116
Report Number1030489-2022-00604
Device Sequence Number20112970
Product Code LXH
UDI-Device Identifier00643169492455
UDI-Public00643169492455
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6642002
Device Catalogue Number6642002
Device Lot NumberEM16K037
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2022
Initial Report FDA Received Date07/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexMale
-
-