Video shows iol(intraocular lens) implantation with company cartridge.Iol and cartridge preparation is not recorded.Iol is advanced at the wound guard, iol position for the advancement appears to be in a correct position.Iol is implanted and a large mark/ line is visible over the iol optic.Based on our observation of the attached video, a large mark/ line is visible over the iol optic.A definitive determination of damage cannot be made without the evaluation of the physical product.A final root cause cannot be determined based on available information.Lens damage may occur: due to the use of a non-qualified viscoelastic.Lens delivery performance may be negatively affected when using other non-qualified viscoelastic leading to underfill, overfill, misfolding , delivery issues and /or damage.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly and cause delivery issues and /or damage.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement and cause delivery issues and / or product damage.Any of the above listed causes alone, or in combination, may create the reported event.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|