MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON TORIC IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON TORIC IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number CNW0T9
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
An ophthalmic surgeon reported that during an intraocular lens (iol) implant procedure, a scratch from the root of the haptic to the front of the center of the optical section was confirmed after implantation.The surgery was completed without product replacement.The sample is not available for return as it was still implanted in the patient's eye.Additional information has been requested and received that the surgeon suspects both iol and cartridge.No health hazard at this time.
 
Additional Manufacturer Narrative
Video shows iol(intraocular lens) implantation with company cartridge.Iol and cartridge preparation is not recorded.Iol is advanced at the wound guard, iol position for the advancement appears to be in a correct position.Iol is implanted and a large mark/ line is visible over the iol optic.Based on our observation of the attached video, a large mark/ line is visible over the iol optic.A definitive determination of damage cannot be made without the evaluation of the physical product.A final root cause cannot be determined based on available information.Lens damage may occur: due to the use of a non-qualified viscoelastic.Lens delivery performance may be negatively affected when using other non-qualified viscoelastic leading to underfill, overfill, misfolding , delivery issues and /or damage.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly and cause delivery issues and /or damage.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement and cause delivery issues and / or product damage.Any of the above listed causes alone, or in combination, may create the reported event.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON TORIC IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14905076
Report Number9612169-2022-00313
Device Sequence Number20093663
Product Code MJP
UDI-Device Identifier00380652412483
UDI-Public00380652412483
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Model NumberCNW0T9
Device Lot Number21284672
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/06/2022
Supplement Date Received by Manufacturer09/02/2022
Initial Report FDA Received Date07/04/2022
Supplement Report FDA Received Date09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INTREPID AUTOSERT IOL HANDPIECE; MONARCH III IOL, CARTRIDGE D; MONARCH III IOL, INJECTOR; PROVISC
Patient SexUnknown
-
-