This report is part of a retrospective review and remediation efforts in response to a warning letter.Correction/removal number: z-0956-2020.(b)(4).S.W version 4.11e.Pump was received with a serial number label fading.The test reservoir does lock properly inside the reservoir tube.Pump powered up properly after battery installation.No blank display noted.Pump passed the displacement test, sleep current measurement, active current measurement and self test.Pump was received with normal operating currents.Pump was monitored for several hours and no unexpected powering off or blank display noted.Pump was also monitored without the test energizer battery inside the battery compartment and reinserts it back into the battery compartment for less than 10 minutes and no unexpected pump error 23 alarm noted.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|