• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON PANOPTIX TORIC TRIFOCAL HYDROPHOBIC IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON PANOPTIX TORIC TRIFOCAL HYDROPHOBIC IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number CNWTT6
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
An ophthalmic surgeon reported that during an intraocular lens (iol) implant procedure, a scratch from the root of the haptic to the front of the center of the optical section was confirmed after implantation.The surgery was completed without product replacement.The sample is not available for return as it was still implanted in the patient's eye.Additional information has been requested and received that the surgeon suspects both iol and cartridge and requested the investigation of cartridge too.No health hazard at this time.
 
Additional Manufacturer Narrative
A video was provided.Video on the lens preparation for the implantation is not provided.The cartridge tip comes in the picture, the lens position for the implantation appears to be acceptable.The haptics appears to be folded correctly.The lens is implanted and a mark/line is observed around the optic edge area, the mark/line appears to be on the posterior surface of the iol.The mark could not be removed/cleaned.Based on our observation of the attached video, a mark/line is visible over the iol optic.A definitive determination of damage cannot be made without the evaluation of the physical product.A final root cause cannot be determined based on available information.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX TORIC TRIFOCAL HYDROPHOBIC IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14907160
Report Number9612169-2022-00314
Device Sequence Number12034479
Product Code MJP
UDI-Device Identifier00380652469999
UDI-Public00380652469999
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2025
Device Model NumberCNWTT6
Device Catalogue NumberCNWTT6.135
Device Lot Number25267420
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/06/2022
Supplement Date Received by Manufacturer07/10/2022
09/02/2022
Initial Report FDA Received Date07/04/2022
Supplement Report FDA Received Date08/08/2022
09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INTREPID AUTOSERT IOL HANDPIECE; MONARCH III IOL, CARTRIDGE D; MONARCH III IOL, INJECTOR; PROVISC
Patient SexUnknown
-
-