• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 32MM X 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 32MM X 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71337650
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a thr surgery was performed, patient underwent a revision surgery on (b)(6) 2022 were the head and liner were exchanged.Current health status of the patient is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after a thr surgery was performed, patient underwent a revision surgery on (b)(6) 2022 in which the r3 20 degree xlpe acetabular liner 32mm inner diameter x outer diameter 50mm was exchanged.Current health status of the patient is unknown.
 
Additional Manufacturer Narrative
Corrected data: b5, d1, d2, d4 (catalog number and udi), h6 (health effect - clinical code).
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, no relevant supporting clinical information has been provided to assist with a clinical investigation.Therefore, based on insufficient information, a thorough clinical assessment cannot be performed at this time.The patient's current condition is unknown and the patient impact beyond the reported events could not be determined based on the limited information provided.Should any additional clinical information be provided this complaint will be re-evaluated.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for total hip systems provides complete guidelines of indications, contraindications, warnings and precautions and possible adverse effects that may occur preoperative, during surgery or post operative.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or some potential probable causes that could contribute to the reported event have been identified as overuse or excessive pressure on the joint, injury, infection and/or patient condition.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 20 DEG XLPE ACET LNR 32MM X 50MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14910942
Report Number1020279-2022-03210
Device Sequence Number5163247
Product Code JDI
UDI-Device Identifier03596010598486
UDI-Public03596010598486
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71337650
Device Catalogue Number71337650
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2022
Supplement Date Received by Manufacturer07/04/2022
08/09/2022
Initial Report FDA Received Date07/04/2022
Supplement Report FDA Received Date07/06/2022
08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient SexUnknown
-
-