It was reported that during an angioplasty procedure in the left common iliac vein, the balloon was allegedly ruptured.It was further reported that the balloon membrane became entangled and wedged in the in the vessel in such a way that even with further deflation, twisting and pulling of the balloon catheter, it could not be removed from the balloon catheter could not be removed from the patient with the usual amount of force.Reportedly, the balloon was removed under massive the balloon membrane detached from the catheter and remained on the guide wire in the patient.It was further reported that the open surgery via common femoral vein was required to remove detached segment from patient.The current status of the patient is unknown.
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one atlas gold pta dilatation catheter in two segments was returned for evaluation.One segment was noted to have a fragmented balloon piece, and another segment consisted of the remaining portion of the catheter and exposed inner guide wire lumen.All the anomalies are noted on microscopic observation.No functional testing was performed due to the condition of the sample and the nature of the complaint.Therefore, the investigation remains inconclusive for the reported balloon rupture as the returned balloon was disintegrated into multiple fragments and returned for evaluation.The investigation also remains inconclusive regarding the reported entrapment as the condition of use could not be replicated under laboratory conditions.However, the investigation is confirmed for the reported balloon detachment as the balloon was noted to have completely detached from the catheter and disintegrated into multiple fragments.A definitive root cause for the reported balloon rupture, entrapment and balloon detachment could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 10/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that during an angioplasty procedure in the left common iliac vein, the pta balloon allegedly ruptured and got stuck in the treated segment of the patient's vein.It was further reported that the catheter had to be pulled back with increased force.Reportedly, the balloon catheter disintegrated with the balloon membrane remaining in the vascular flow path.The detached segments were removed by open surgery.The current status of the patient is unknown.
|