• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FINELINE II EZ STEROX; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FINELINE II EZ STEROX; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 4474
Medical Device Problem Codes Fracture (1260); Difficult to Remove (1528)
Health Effect - Clinical Codes Unspecified Infection (1930); Foreign Body In Patient (2687)
Date of Event 06/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient with this right ventricular (rv) lead had an infection.A revision was performed and the pacemaker was explanted while the right ventricular (rv) lead and right atrial (ra) lead were capped.In addition, the tip of the leads were retained in the heart following the procedure.The physician fractured the leads in an attempt to extract them.No additional adverse patient effects were reported.
 
Event or Problem Description
It was reported that the patient with this right ventricular (rv) lead had an infection.A revision was performed and the pacemaker was explanted while the right ventricular (rv) lead and right atrial (ra) lead were capped.In addition, the tip of the leads were retained in the heart following the procedure.The physician fractured the leads in an attempt to extract them.Multiple attempts were made to get all parts of the lead, but were unsuccessful.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.Correction to field b5: describe event or problem, f10: device codes and h6: device codes.
 
Event or Problem Description
It was reported that the patient with this right ventricular (rv) lead had an infection.A revision was performed and the pacemaker was explanted while the right ventricular (rv) lead and right atrial (ra) lead were capped.In addition, the tip of the leads were retained in the heart following the procedure.The physician fractured the leads in an attempt to extract them.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.Correction to field b5: describe event or problem, f10: device codes and h6: device codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FINELINE II EZ STEROX
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
#12, rd. #698
dorado PR 00646
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14913779
Report Number2124215-2022-24039
Device Sequence Number2031677
Product Code DTB
UDI-Device Identifier00802526265037
UDI-Public00802526265037
Combination Product (Y/N)Y
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P960004/S014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2004
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/01/2004
Device Model Number4474
Device Catalogue Number4474
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/20/2022
Supplement Date Received by Manufacturer06/20/2022
07/28/2022
Initial Report FDA Received Date07/05/2022
Supplement Report FDA Received Date07/07/2022
08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/01/2002
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age66 YR
Patient SexMale
-
-