MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. PROXIMATE*CUTTER RELOAD UNIT; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. PROXIMATE*CUTTER RELOAD UNIT; STAPLE, IMPLANTABLE Back to Search Results
Model Number TCR10
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date of event unknown, captured as awareness date.Batch # r59y7c.Investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned reload.Visual analysis of the returned sample determined that one tcr10 reload was received with no apparent damage.The reload was received with 24 drivers up without staples scattered along the staple line.In addition, the swing tab was noted to be detached and missing, making the reload nonfunctional.Due to the condition of the reload no functional test was performed to preserve the evidence.No conclusion could be reached on what caused the swing tab to detach and the drivers up, it should be noted that each reloads is 100% inspected through an automated vision system to ensure that the swing tab and staples are present and in the correct position prior to shipment.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
 
Event or Problem Description
The device received has been classified as an unreconciled blind unit.No further information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
4188 lexington avenue north
shoreview 55126
MDR Report Key14914180
Report Number3005075853-2022-04325
Device Sequence Number16651986
Product Code GDW
UDI-Device Identifier10705036003045
UDI-Public10705036003045
Combination Product (Y/N)N
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/30/2023
Device Model NumberTCR10
Device Catalogue NumberTCR10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/15/2022
Initial Report FDA Received Date07/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/11/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-