According to the complaint, on (b)(6) 2022 a sample was measured on the abl90 flex analyzer (serial number: (b)(4)) and the following measurements were obtained on the sodium (na) parameter: 21:38 138 mmol/l, 21:39 121 mmol/l, 22:22 107 mmol/l, 22:24 107 mmol/l.A control ionogram was conducted with the sample and the following measurement was obtained on the sodium (na) parameter: 137 mmol/l (comparison measurement).Based on these measurements, the customer reported the last 3 results from the abl90 flex analyzer as false low.
|