• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SUTURE ANCHOR, MICRO CORKSCREW FT; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SUTURE ANCHOR, MICRO CORKSCREW FT; SCREW, FIXATION, BONE Back to Search Results
Model Number SUTURE ANCHOR, MICRO CORKSCREW FT
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 06/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On 6/14/2022 it was reported by a sales representative via sems that an (b)(4) suture anchor broke during a ulnar sheath repair.The implant broke during insertion where the driver meets the anchor and the sutures were frayed.All was retrieved.A second suture anchor was used and that one also frayed the sutures during insertion.All debris was retrieved and the procedure was completed successfully.No further information provided.Additional information: they were unable to remove the second defective anchor from the patient.It is countersunk in the bone but will not cause irritation.To complete the case a larger anchor was used (micro corkscrew) in the same drill hole as an alternative.
 
Additional Manufacturer Narrative
Complaint confirmed.Upon visual inspection, it was noted that the anchor was not returned for investigation.It was also noted that the sutures were broken near where the anchor would be located.The plastic clip has been deployed and the tip where the anchor is seated is notably twisted.The method used to prepare the bone as well as the bone quality encountered were not provided.The root cause remains undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTURE ANCHOR, MICRO CORKSCREW FT
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key14914522
Report Number1220246-2022-05154
Device Sequence Number20377482
Product Code HWC
UDI-Device Identifier00888867008519
UDI-Public00888867008519
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K112237
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSUTURE ANCHOR, MICRO CORKSCREW FT
Device Catalogue NumberAR-1318FT
Device Lot Number14749170
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2022
Supplement Date Received by Manufacturer06/14/2022
Initial Report FDA Received Date07/05/2022
Supplement Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-