|
Model Number FIBERTAK |
Device Problems
Device Slipped (1584); Output Problem (3005)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 05/03/2022 |
Event Type
malfunction
|
Event Description
|
On 05/04/2022, it was reported by a sales representative via phone that (3) ar-3638 fibertaks failed.This occurred on (b)(6) 2022 during a bankart repair when the surgeon cycled the drill bit too many times resulting in the implant backing out.A second anchor was implanted properly into the same hole however it cinched up 3/4 of the way, and would not lock.The anchor remained implanted, and the sutures were cut out.A third implant was used and cycled too many times and resulted in the device backing out.The case was completed using 3 more fibertaks that were not cycled more than they should be and were implanted successfully.There was no patient effect reported.Additional information provided on 05/04/2022 that the 1.8 mm solid drill bit from the perc kit that was being used.The patient had a little softer bone quality.During returned device evaluation, a reportable malfunction was discovered.
|
|
Manufacturer Narrative
|
During returned device evaluation, a reportable malfunction was discovered.The complaint is unable to be confirmed.Two unpackaged ar-3638 inserters were received for investigation.Visual inspection identified that neither of the two c16235-10 suture tape assemblies corresponding to the returned inserters were returned for analysis.As such, the implant failure allegations cannot be verified.It was noted that one of the two prongs at the distal tip of one of the inserters had broken off.No fragments were returned.The cause remains undetermined.The method used to prepare the bone, as well as the bone quality encountered, were not provided.The thickness of the broken inserter tip was assessed using digital micrometer id: 224, and was found to meet design specifications.
|
|
Search Alerts/Recalls
|
|
|