• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. FASTPASS SCORPION-MULTIFIRE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. FASTPASS SCORPION-MULTIFIRE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number FASTPASS SCORPION-MULTIFIRE
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
On 6/21/2022, it was reported by an arthrex employee via sems that an ar-13997mf multifire fastpass scorpion head broke.This was discovered during an rcr procedure on (b)(6) 2022.Surgeon used a suture lasso to complete the case.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Complaint confirmed.Upon visual inspection, it was noted that the upper jaw of the device had broken at the trapdoor.The trap door and broken piece of the upper jaw were not returned for investigation.The cause of this remains undetermined however, probable cause can be attributed to the application of prying/leveraging forces.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FASTPASS SCORPION-MULTIFIRE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key14915117
Report Number1220246-2022-05181
Device Sequence Number2033272
Product Code LXH
UDI-Device Identifier00888867015975
UDI-Public00888867015975
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFASTPASS SCORPION-MULTIFIRE
Device Catalogue NumberAR-13997MF
Device Lot Number10235413
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2022
Supplement Date Received by Manufacturer06/14/2022
Initial Report FDA Received Date07/05/2022
Supplement Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/31/2018
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-