| Model Number |
2901-N90098-B |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the tip of an everest locking screw inserter became deformed intra-operatively.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned everest navigated locking screw inserter was visually inspected.The device appeared cosmetically worn from normal use.The tip was confirmed to be damaged; majority of the hexalobe had fractured off, leaving a relatively even fracture surface.The fracture surface revealed a rotational pattern, indicating the device failed during torsion.Device history records were reviewed and no relevant manufacturing issues were identified.Complaint history was reviewed and no similar complaints for the reported lot number were identified.The most likely root cause for the tip fracture can be attributed to normal device wear over time, as the instrument has been in the field for approximately 3 years.Consistent use of the instrument throughout its lifetime may cause material fatigue, leading to eventual tip fracture.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of an everest locking screw inserter became deformed intra-operatively.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.Inspection of the returned device identified tip fracture rather than tip deformation.
|
| |
|