• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ANGIOJET ULTRA 5000A; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ANGIOJET ULTRA 5000A; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 45038
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was cancelled.An angiojet ultra 5000a was advanced for treatment.During the procedure, the unit shut off automatically and a system error code 64 occurred.The device were unable to troubleshoot the issue.The procedure was cancelled.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: angiojet console was returned.A visual inspection was done on the console, and it failed for multiple reasons.Cracked display, broken hook left side, handle dented, foot pedal holder broken off, power cord missing, thread insert broken off of tower main top, and also thick dust all over inside of panels and on drawer switch.Powered on the unit, and the complaint was confirmed as the unit would not power up.Removed and replaced with a golden display; it then powered on as expected.
 
Event or Problem Description
It was reported that the procedure was cancelled.An angiojet ultra 5000a was advanced for treatment.During the procedure, the unit shut off automatically and a system error code 64 occurred.The device were unable to troubleshoot the issue.The procedure was cancelled.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET ULTRA 5000A
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14915336
Report Number2134265-2022-07593
Device Sequence Number5813560
Product Code MCX
UDI-Device Identifier08714729890584
UDI-Public08714729890584
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P980037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/07/2025
Device Model Number45038
Device Catalogue Number45038
Device Lot NumberOEMTRACE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/11/2022
Supplement Date Received by Manufacturer11/04/2022
Initial Report FDA Received Date07/05/2022
Supplement Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/24/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-