• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 14MM X 50CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 14MM X 50CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0036121450
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the procedure, the subject coil was found prematurely deployed inside the microcatheter.No further information available and there was no clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during the procedure, the subject coil was found prematurely deployed inside the microcatheter.No further information available and there was no clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
D4 ¿ lot# - corrected.Although the lot number was not provided, the automated manufacturing execution system (mes) has controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, visual and functional testing as well as physical analysis cannot be performed.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.Additional information provided by the customer indicated that coils prematurely detached in catheter only and one was entirely retrievable.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable was assigned to this complaint.H3 other text : the device is not available to the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XL 360 SOFT 14MM X 50CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14916730
Report Number3008881809-2022-00325
Device Sequence Number1932659
Product Code HCG
UDI-Device Identifier04546540711540
UDI-Public04546540711540
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM0036121450
Device Catalogue NumberM0036121450
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2022
Supplement Date Received by Manufacturer08/30/2022
Initial Report FDA Received Date07/05/2022
Supplement Report FDA Received Date09/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MICROCATHETER (UNKNOWN); TARGET XL 360 SOFT EMBOLIZATION COIL (STRYKER)
Patient SexUnknown
-
-