DEPUY MITEK LLC US 5.0 MM BARREL BURR PLUS; ARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE Back to Search Results
Model Number 283889
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 06/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
This is report 2 of 2 for (b)(4).It was reported by the affiliate in that during a shoulder decompression and arthroscopic acromio-clavicular (acj) excision, it was observed that there were metal filings left in the joint while using the 5.0 mm barrel burr plus blade device.There were no adverse patient consequences nor surgical delay reported.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: during subsequent follow-up with the customer, additional information was obtained.The reporter stated that there was a 5-minute delay in the procedure due to the event.It was noted that all the metal shavings were not removed from the patient.There was no additional medical/surgical intervention required to treat the issue, and the patient¿s current status is unknown.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H6: the health effect - impact code has been updated accordingly.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.However, photos were provided.Upon visual inspection of the photos, it appeared that the inner had signs of friction and striation marks on the surface of the distal may cause metal loss.A manufacturing record evaluation was performed for the finished device m2107010, and no non-conformances were identified.Based on device condition observed in the photo, this complaint can be confirmed.No further procedure information was provided to determine how this failure occurred.The photos do not provide enough evidence to determine root cause.The possible root cause for reported failure can be attributed to blade wear and degradation.As per ifu; excessive side loading may result in blade wear and degradation.Adequate suction is recommended to reduce wear and degradation of the device.Besides, the handpiece where the blade is assembled was not received for evaluation, potential ways to reduce shedding is to ensure all hand pieces are properly serviced and maintained.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthese mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0 MM BARREL BURR PLUS
Common Device Name
ARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key14917466
Report Number1221934-2022-02006
Device Sequence Number2031777
Product Code HRX
UDI-Device Identifier10886705022106
UDI-Public10886705022106
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K131191
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model Number283889
Device Catalogue Number283889
Device Lot NumberM1708021
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2022
Supplement Date Received by Manufacturer07/21/2022
07/22/2022
08/16/2022
Initial Report FDA Received Date07/05/2022
Supplement Report FDA Received Date07/21/2022
07/22/2022
08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/11/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-