| Model Number |
TV-AB2080-E |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
06/15/2022
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
The devices involved in this event will not be returned for evaluation.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device not available for return.
|
| |
|
Event or Problem Description
|
|
The patient was initially treated for an abdominal aortic aneurysm (aaa) with the implant of the ovation prime abdominal aortic aneurysm stent system.Approximately seven (7) years post initial procedure, the patient presented with abdominal pain in the flank and iliac fossa.The patient had an explant of the device due to an infection possibly caused by the patient contracting malaria.The final patient status was reported as being alive and well.
|
| |
|
Event or Problem Description
|
|
The patient was initially treated for an abdominal aortic aneurysm (aaa) with the implant of the ovation prime abdominal aortic aneurysm stent system.Approximately seven (7) years post initial procedure, the patient presented with abdominal pain in the flank and iliac fossa.The patient had an explant of the device due to an infection possibly caused by the patient contracting malaria.The final patient status was reported as being alive and well.After the initial report, the clinical evaluation also shows reasonable evidence to suggest indicating the patient passed away on (b)(6) 2022 due to the infected graft.The patient death was discovered during a review of the discharge/death summary dated (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because the device was explanted and not available for return.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the ovation prime graft infection with explant is confirmed.This is consistent with the reported adverse event/incident.The clinical evaluation also shows reasonable evidence to suggest the patient passed away on (b)(6) 2022.These findings were discovered during an examination of the discharge/death summary dated (b)(6) 2022.It was reported the patient was transferred from nicaragua with a positive malaria antigen.It is unclear if the malaria was the source of the graft infection.Blood cultures were positive on (b)(6) 2022 for burkholderia cepacia complex.The most likely causation for the patient death was determined to be device related (infected graft).The procedure related harms identified were abnormal blood loss, hemodynamic instability, mesenteric ischemia with need for additional surgical procedure, acute renal failure, and death.The final patient status was reported to be expired post operative day two, hospital day twenty-one.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.H6: investigation finding codes - remove code 3233 h6: investigation conclusion codes - remove code 11.
|
| |
|