• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH LATX FR 14FR POVIDONE IOD 10OZ; TRAY, CATHETERIZATION, STERILE URETHRAL, WITH OR WITHOUT CATHETER (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH LATX FR 14FR POVIDONE IOD 10OZ; TRAY, CATHETERIZATION, STERILE URETHRAL, WITH OR WITHOUT CATHETER (KIT) Back to Search Results
Model Number 3450LF
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
Customer reports: there were multiple occurrences with one patient where the devices were not working.For instance, two device kits were used with no urine return, but with a third try was successful.The next time this patient needed to be catheterized, the same thing happened.The customer has pulled all kits that had the same lot number as the affected products.
 
Additional Manufacturer Narrative
The device was received for evaluation and the reported condition has been confirmed.  a corrective and preventive action has been initiated to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATX FR 14FR POVIDONE IOD 10OZ
Common Device Name
TRAY, CATHETERIZATION, STERILE URETHRAL, WITH OR WITHOUT CATHETER (KIT)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 22500
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key14919595
Report Number9612030-2022-03306
Device Sequence Number2032467
Product Code FCM
UDI-Device Identifier10884521009585
UDI-Public10884521009585
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number3450LF
Device Catalogue Number3450LF
Device Lot Number2210219764
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/29/2022
Supplement Date Received by Manufacturer06/29/2022
Initial Report FDA Received Date07/05/2022
Supplement Report FDA Received Date08/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-