Model Number ZCU450
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Disturbances (2140)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the intraocular lens (iol), model zcu450 with 15.0 diopter power, was implanted in the patient¿s right eye.The iol was explanted in a separate surgical procedure and replaced with a zcu150 model lens with 15.5 diopter power due to visual impairment that was first identified during post-operative examination.It was reported that the patient was temporarily impaired.Account indicated that there were no complications, such as capsular tear, no incision enlargement, no serious patient injury, no unplanned suture(s), and no vitrectomy during the iol explant procedure.Patient outcome post-lens exchange is unknown.It is also unknown if the patient is still temporarily impaired post-lens exchange.No further information was provided.
 
Additional Manufacturer Narrative
Additional information: patient weight: unknown as information was asked but not provided.Date of event: unknown as information was not provided; the best estimate date is between 31-mar-2022 and 15-jun-2022.The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts were made to contact the customer account requesting additional information regarding the complaint; however, to date no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Product investigation was completed, and the following fields were updated accordingly: section d-9: device available for evaluation: yes.Section d-9: date returned to manufacturer: 18-aug-2022.Section h-3: device returned to manufacturer: yes.Device evaluation: the complaint lens was received in an unknown solution inside a specimen cup.Visual inspection under magnification revealed the lens was returned cut in half.No defects that could contribute to visual issues were observed.No additional defects were observed.Complaint issue ¿visual disturbance¿ was not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: based on the investigation results, there is no indication of a product malfunction or product quality deficiency.The reported issue was not verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14921580
Report Number3012236936-2022-01687
Device Sequence Number19330165
Product Code HQL
UDI-Device Identifier05050474663008
UDI-Public(01)05050474663008(17)240924
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/24/2024
Device Model NumberZCU450
Device Catalogue NumberZCU450U150
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2022
Supplement Date Received by Manufacturer09/21/2022
Initial Report FDA Received Date07/05/2022
Supplement Report FDA Received Date10/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/24/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-