SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102812-81A
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/22/2022
Event Type  malfunction  
Event Description
On (b)(6) 2022, senseonics was made aware of an event where the user experienced sensor suspend alert leading to early sensor removal.
 
Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Manufacturer Narrative
The customer's complaint was confirmed during in-vitro testing and during the review of in-vivo data.The sensor triggered 7 sensor suspend alarms as it was observed that the system had rejected 4 consecutive calibrations and the led experienced a permanent crash (confirmed throughout the whole qc test).The system's self-test functions worked normally by disabling the sensor due to the detected performance failure.As part of resolution, the rma was authorized to offer the user a sensor replacement.A new transmitter was sent to the user as well, under rma-18351 (complaint: (b)(4).D9 device available for evaluation? yes, device received on 10 august 2022.H3.Device evaluated by manufacturer? yes.H6.Type of investigation updated to 10 h6.Investigation findings updated to 4203 h6.Investigation conclusions updated to 4307.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20875-7005
MDR Report Key14922473
MDR Text Key295345361
Report Number3009862700-2022-00111
Device Sequence Number1
Product Code QHJ
UDI-Device Identifier00817491022974
UDI-Public817491022974
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/17/2023
Device Model Number102812-81A
Device Catalogue NumberFG-7202-00-301
Device Lot NumberWP08816
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received06/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/14/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age52 YR
Patient SexFemale
-
-