The user facility reported that the wire became unraveled while being used with cook cxi support cath during a procedure.There was no patient injury, and no medical or surgical intervention was required.The patient was in stable condition.The procedure outcome was unknown.The event occurred intra-operative.Additional information was received 17 jun 2022: an angiogram of the leg was performed.The wire was stuck inside an.014 cxi catheter.They were able to pull it out, however the coating began to come off and was no longer safe to put back inside the patient's body.The procedure was completed successfully, without opening another wire.Another.014 glide advantage was used to cross the lesion.Nothing was left inside the patient.
|
|
This report is being submitted as follow-up no.1 to provide the device return date in section d9, to update section h3, and to provide the completed investigation results.Visual inspection of the actual sample found that the distal end had been elongated and the urethane coating had been peeled off.The total length of the actual sample was measured.In comparison with a product of the involved product code, the actual sample had been elongated by approximately 20 mm.The actual sample total length was approximately 3020mm.The product of the involved product code total length was approximately 3000mm.Magnifying inspection of the distal end of the actual sample obtained the following results an elongation of the gold coil was observed and there was a rip approximately 1mm from the distal end.Magnifying inspection of the part where urethane coating had been peeled off obtained the following result a streaky scratch was and was confirmed that the urethane coating was peeling off toward the distal end.Magnifying inspection of the ptfe coating section found peeling of the ptfe coating and scratches on the part from approximately 310mm to 480mm from the distal end.Electron microscopic inspection of the distal end section found no flaws in the area other than where the rip was found under a magnifier.Electron microscopic inspection of the part in which the urethane coating had been peeled off found that there were streaky scratches on the core wire.Electron microscopic inspection of the ptfe coating section found a shape that looked like the ptfe coating has been torn off.X-ray fluoroscopic inspection of the distal end obtained the following results the distal end of the core wire was located at approximately 20mm from the distal end of the actual sample and the fracture of the core wire was not observed.The outer diameter of the actual sample was measured and confirmed to meet the factory's specifications.A search of the complaint file found no other similar report with the involved product code/lot number combination from other facilities.In the past, the following simulation test were conducted.Based on an assumption that a guidewire was bent and was pulled out in a state of strong contact with the metal device, a guide wire was pulled out from a metal device.As a result, peeling of the ptfe coating was observed.Ashitaka factory manufacturing process assures the quality of this product by performing 100% visual inspection to confirm that the wire is not exposed and there is no anomaly in the appearance.Ifu (instructions for use) of this product includes the following statements: (i) if any resistance is felt or if the distal end's behavior and/or location seems improper, stop manipulating the glidewire advantage and/or the catheter and determine the cause by fluoroscopy.Continuing to manipulate or rotate the glidewire advantage or failure to exercise proper caution may result in bending, kinking, separation of the guide wire's distal end, damage to the catheter, or damage to the vessel.(ii) do not use the glidewire advantage with device which contain metal parts such as atherectomy catheters, laser catheters, or metal introduction device as they may cause the glidewire advantage plastic coating to shear and/or sever the wire.
|