Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during surgery the drill fractured inside the patient¿s bone.The piece could not be removed and remained inside the patient.Unknown impact or consequence to the patient.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Additional Manufacturer Narrative
(b)(4).Source foreign: (b)(6).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and corrected information.The complained product was returned for examination and the reported event can be confirmed.A large portion of the drill bit has fractured off.Review of the device history records identified no deviations or anomalies during manufacturing.The raw material certificate was reviewed with no anomalies noted.Therefore, the investigation did not identify a nonconformance or a complaint out of box (coob).Device is used for treatment.Medical records were not provided.Although the instrument is a reusable instrument, the number of instrument usage is unknown.It is likely that a high application of force or a bad drilling angle caused or contributed to the reported event.However, based on the received information and details of the investigation, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWO-FLUTED DRILL BIT 2.5 MM, WITH QUICK COUPLING, L. 154/180 MM
Common Device Name
TRAUMA INSTRUMENT
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14926780
Report Number0009613350-2022-00346
Device Sequence Number12313054
Product Code HTW
UDI-Device Identifier00889024298064
UDI-Public(01)00889024298064(11)130226(10)PF13958
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number10325180
Device Lot NumberPF13958
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2022
Supplement Date Received by Manufacturer08/12/2022
Initial Report FDA Received Date07/06/2022
Supplement Report FDA Received Date08/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/26/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-